Gap
Which clinical evidence is missing for a specific claim or risk.
MDR · Articles 62–82 · Annex XV
An investigation should answer a predefined clinical question, not simply produce more data.
The regulatory pathway, design, risk management and analysis should form one traceable system from the clinical-evaluation gap to the final report.
Starting point
A clinical investigation is justified when the available data cannot reliably demonstrate safety, clinical performance or an acceptable benefit-risk ratio for the intended purpose and target population.
The gap should be defined in the clinical evaluation. It then determines the objective, hypothesis, endpoints, population, follow-up period and analysis plan. If that link is lost, even a technically sound investigation may fail to answer the notified body's or competent authority's question.
Which clinical evidence is missing for a specific claim or risk.
In whom, with which device, compared with what and over which period the outcome is assessed.
How bias will be limited and interpretable data obtained.
How the result will change the clinical evaluation, risks, labelling or further PMCF.
Regulatory pathway
The pathway depends on the investigation's purpose, CE status, use within the intended purpose and the nature of additional procedures. The sponsor is responsible for selecting the correct regulatory route.
Conformity assessment
Investigations conducted to demonstrate conformity with the MDR, including establishing or verifying safety, clinical performance and clinical benefit.
Articles 62–81 and Annex XV apply.After CE marking
For a CE-marked device used within its intended purpose, Article 74 establishes a specific route when subjects undergo additional invasive or burdensome procedures.
Use outside the intended purpose brings the investigation back under Articles 62–81.Other purposes
Investigations not conducted for the purposes listed in Article 62(1) must meet the core protections of Articles 62(2) and 62(3), as well as applicable national requirements.
The specific procedure varies between Member States.IVDs follow different terminology and a different legal regime — performance studies under Regulation (EU) 2017/746. They should not automatically be described as clinical investigations under the MDR.
Qualification
This map does not replace a legal assessment, but it shows why one factor — such as the presence of a CE mark — is not enough to select the procedure.
The safety or clinical performance of a specific medical device.
For initial conformity assessment, PMCF or another scientific purpose.
Within its intended purpose, outside it or before CE marking.
Invasive, burdensome or other procedures beyond normal care.
Before the first subject
The clinical evaluation, risk management, CIP, Investigator's Brochure, informed-consent documents and statistical plan are not independent files. Differences in intended purpose, population, device version or risks create a regulatory and ethical problem.
Rationale for the investigation, MDR pathway, countries, sites and the role of the results in conformity assessment.
The CIP covering objectives, design, population, endpoints, monitoring, data handling and the method of analysis.
The current body of preclinical and clinical data, device characteristics and risk information.
A clear description of the investigation, procedures, alternatives, risks, rights, data protection and consent process.
Submission forms, insurance, site qualification, contracts, monitoring, safety, data management and version control.
Good clinical practice
The fourth edition of ISO 14155 was published in March 2026. It covers the design, conduct, recording and reporting of medical device investigations involving human subjects, as well as sponsor and principal-investigator responsibilities.
Open the ISO 14155:2026 pageThe subject's interests take priority, and consent is a process rather than only a signature.
Investigation risks are justified only when the design can produce a reliable answer.
Device, procedural, use-error and investigation-conduct risks all require control.
Decisions, data, deviations, events and document versions must remain verifiable.
Important clarification
Medical device software is not excluded from clinical investigations merely because it does not physically contact the patient. ISO expressly allows the standard to be applied to Software as a Medical Device.
The standard does not apply directly to IVDs. Performance studies are governed by the IVDR, although individual GCP principles may be useful where the applicable rules permit.
Conduct
The sponsor must maintain a state of control throughout the investigation, including work performed by contractors. Transferring tasks to a CRO does not transfer the sponsor's responsibility.
Investigator qualification, training, contracts, equipment, document versions and investigational-device handling.
Risk-based oversight of CIP compliance, subject protection, source data and critical processes.
Timely assessment and reporting of adverse events, serious adverse events and device deficiencies.
Management of queries, deviations, changes, missing data, database lock and the audit trail.
After the last visit
Finalise the analysis populations, address deviations and missing data, and perform the prespecified analysis.
The clinical investigation report should present methods and results transparently, including limitations and unfavourable findings.
Under MDR Article 77, the report and summary are submitted within the prescribed period, with a shorter period after early termination.
Results feed into the CER, risk-management file, PMS and PMCF, instructions, claims and further strategy.
EUDAMED and national procedures
As of October 2026, the EUDAMED Clinical Investigations and Performance Studies module is not yet mandatory and is identified by the European Commission as still under development. Existing national procedures and transitional arrangements apply until its mandatory use begins.
For a multicentre investigation across several Member States, one package cannot be assumed to satisfy every national requirement automatically. Document language, ethics review, insurance, timelines and local forms should be checked for every country in advance.
European Commission sectionPrimary sources
Articles 61–82 and Annex XV
Questions and answers on clinical investigations
Clinical Investigation Plan content
Investigator's Brochure content
Forms, safety reporting, modifications and report publication
Questions and answers
No. The need for an investigation depends on the clinical evaluation, the device class and characteristics, the available clinical data and the requirements of MDR Article 61. Specific requirements and limited exemptions apply to implantable and class III devices.
Yes. The regulatory pathway depends on the investigation's purpose, whether the device is used within or outside its intended purpose, and whether additional invasive or burdensome procedures are involved. PMCF investigations within the intended purpose may fall under MDR Article 74.
Yes, when the software is a medical device and the investigation involving human subjects assesses its clinical performance, effectiveness or safety. The absence of physical contact with a patient does not by itself exclude the principles of good clinical practice.
No. IVD performance studies are governed by the IVDR. Individual ISO 14155 principles may be useful where the applicable rules permit, but the regulatory pathway and documentation are determined by the IVDR.
Ethics committee approval is not sufficient on its own. Before the investigation starts, the applicable competent-authority and national requirements must be met, the required opinions and authorisations obtained, contracts and insurance completed, sites prepared, and the documentation and device confirmed as ready.
A sponsor may transfer individual tasks to a CRO but retains responsibility for proper investigation management and oversight of the transferred functions. The scope of delegation, interfaces, quality control and escalation should be documented.
Before writing the CIP
Describe the device, its class and intended purpose, the current clinical evaluation, the proposed population and the question the investigation must answer. MDRC can connect the regulatory pathway, design and documentation package.