Clinical investigations of medical devices in the EU

An investigation should answer a predefined clinical question, not simply produce more data.

The regulatory pathway, design, risk management and analysis should form one traceable system from the clinical-evaluation gap to the final report.

Legal basisMDR 2017/745 GCPISO 14155:2026 OutcomeCredible clinical data

Question first, design second

A clinical investigation is justified when the available data cannot reliably demonstrate safety, clinical performance or an acceptable benefit-risk ratio for the intended purpose and target population.

The gap should be defined in the clinical evaluation. It then determines the objective, hypothesis, endpoints, population, follow-up period and analysis plan. If that link is lost, even a technically sound investigation may fail to answer the notified body's or competent authority's question.

01

Gap

Which clinical evidence is missing for a specific claim or risk.

02

Question

In whom, with which device, compared with what and over which period the outcome is assessed.

03

Design

How bias will be limited and interpretable data obtained.

04

Decision

How the result will change the clinical evaluation, risks, labelling or further PMCF.

Not every investigation follows the same procedure

The pathway depends on the investigation's purpose, CE status, use within the intended purpose and the nature of additional procedures. The sponsor is responsible for selecting the correct regulatory route.

62

Conformity assessment

Article 62 investigations

Investigations conducted to demonstrate conformity with the MDR, including establishing or verifying safety, clinical performance and clinical benefit.

Articles 62–81 and Annex XV apply.
74

After CE marking

PMCF under Article 74

For a CE-marked device used within its intended purpose, Article 74 establishes a specific route when subjects undergo additional invasive or burdensome procedures.

Use outside the intended purpose brings the investigation back under Articles 62–81.
82

Other purposes

Other investigations

Investigations not conducted for the purposes listed in Article 62(1) must meet the core protections of Articles 62(2) and 62(3), as well as applicable national requirements.

The specific procedure varies between Member States.

IVDs follow different terminology and a different legal regime — performance studies under Regulation (EU) 2017/746. They should not automatically be described as clinical investigations under the MDR.

Four questions before preparing the submission

This map does not replace a legal assessment, but it shows why one factor — such as the presence of a CE mark — is not enough to select the procedure.

01

What is assessed

The safety or clinical performance of a specific medical device.

02

Why data are collected

For initial conformity assessment, PMCF or another scientific purpose.

03

How the device is used

Within its intended purpose, outside it or before CE marking.

04

What the protocol adds

Invasive, burdensome or other procedures beyond normal care.

The submission should tell one consistent story

The clinical evaluation, risk management, CIP, Investigator's Brochure, informed-consent documents and statistical plan are not independent files. Differences in intended purpose, population, device version or risks create a regulatory and ethical problem.

01

Clinical and regulatory strategy

Rationale for the investigation, MDR pathway, countries, sites and the role of the results in conformity assessment.

02

Clinical Investigation Plan

The CIP covering objectives, design, population, endpoints, monitoring, data handling and the method of analysis.

03

Investigator's Brochure

The current body of preclinical and clinical data, device characteristics and risk information.

04

Subject information

A clear description of the investigation, procedures, alternatives, risks, rights, data protection and consent process.

05

Application and operational package

Submission forms, insurance, site qualification, contracts, monitoring, safety, data management and version control.

ISO 14155:2026 links subject protection with data credibility

The fourth edition of ISO 14155 was published in March 2026. It covers the design, conduct, recording and reporting of medical device investigations involving human subjects, as well as sponsor and principal-investigator responsibilities.

Open the ISO 14155:2026 page
01

Rights and well-being

The subject's interests take priority, and consent is a process rather than only a signature.

02

Scientific validity

Investigation risks are justified only when the design can produce a reliable answer.

03

Risk control

Device, procedural, use-error and investigation-conduct risks all require control.

04

Traceability

Decisions, data, deviations, events and document versions must remain verifiable.

Important clarification

Medical device software is not excluded from clinical investigations merely because it does not physically contact the patient. ISO expressly allows the standard to be applied to Software as a Medical Device.

The standard does not apply directly to IVDs. Performance studies are governed by the IVDR, although individual GCP principles may be useful where the applicable rules permit.

Protocol approval begins control rather than completing it

The sponsor must maintain a state of control throughout the investigation, including work performed by contractors. Transferring tasks to a CRO does not transfer the sponsor's responsibility.

Site readiness

Investigator qualification, training, contracts, equipment, document versions and investigational-device handling.

Monitoring

Risk-based oversight of CIP compliance, subject protection, source data and critical processes.

Safety

Timely assessment and reporting of adverse events, serious adverse events and device deficiencies.

Data integrity

Management of queries, deviations, changes, missing data, database lock and the audit trail.

The result must return to the evidence system

01

Complete and analyse

Finalise the analysis populations, address deviations and missing data, and perform the prespecified analysis.

02

Prepare the report and summary

The clinical investigation report should present methods and results transparently, including limitations and unfavourable findings.

03

Complete regulatory reporting

Under MDR Article 77, the report and summary are submitted within the prescribed period, with a shorter period after early termination.

04

Update connected documents

Results feed into the CER, risk-management file, PMS and PMCF, instructions, claims and further strategy.

The submission process remains transitional

As of October 2026, the EUDAMED Clinical Investigations and Performance Studies module is not yet mandatory and is identified by the European Commission as still under development. Existing national procedures and transitional arrangements apply until its mandatory use begins.

For a multicentre investigation across several Member States, one package cannot be assumed to satisfy every national requirement automatically. Document language, ethics review, insurance, timelines and local forms should be checked for every country in advance.

European Commission section

Clinical investigations under the MDR

Is a clinical investigation required for every medical device?

No. The need for an investigation depends on the clinical evaluation, the device class and characteristics, the available clinical data and the requirements of MDR Article 61. Specific requirements and limited exemptions apply to implantable and class III devices.

Can a CE-marked device be investigated?

Yes. The regulatory pathway depends on the investigation's purpose, whether the device is used within or outside its intended purpose, and whether additional invasive or burdensome procedures are involved. PMCF investigations within the intended purpose may fall under MDR Article 74.

Does ISO 14155 apply to medical device software?

Yes, when the software is a medical device and the investigation involving human subjects assesses its clinical performance, effectiveness or safety. The absence of physical contact with a patient does not by itself exclude the principles of good clinical practice.

Does ISO 14155 apply directly to IVDs?

No. IVD performance studies are governed by the IVDR. Individual ISO 14155 principles may be useful where the applicable rules permit, but the regulatory pathway and documentation are determined by the IVDR.

Can the investigation start after ethics committee approval?

Ethics committee approval is not sufficient on its own. Before the investigation starts, the applicable competent-authority and national requirements must be met, the required opinions and authorisations obtained, contracts and insurance completed, sites prepared, and the documentation and device confirmed as ready.

Who is responsible when a CRO is involved?

A sponsor may transfer individual tasks to a CRO but retains responsibility for proper investigation management and oversight of the transferred functions. The scope of delegation, interfaces, quality control and escalation should be documented.

Confirm that the design will close the clinical evidence gap

Describe the device, its class and intended purpose, the current clinical evaluation, the proposed population and the question the investigation must answer. MDRC can connect the regulatory pathway, design and documentation package.