A practical classification casebook

One device. Two risk-class systems

An EU medical device class cannot be converted into an FDA class. The product remains the same, but its regulatory logic and supporting rationale must be built separately for each market.

EU22 MDR rules USType and product code OutputTwo rationales

Classify the defined device, not its name

In the EU, the manufacturer applies the definitions and every relevant Annex VIII rule to the stated intended purpose, duration of use, invasiveness, active nature and other characteristics. If several rules apply, the strictest resulting class prevails.

In the US, the device is matched to a legally classified generic device type. The result identifies the regulation number, product code, class and applicable premarket pathway. A predicate becomes relevant only afterwards when the 510(k) route applies.

The same input does not produce the same output

Use one controlled product specification, but do not combine the two regulatory analyses into a device-class conversion table.

EU MDR

Rule, class and conformity-assessment route

01

Define the intended purpose, user, anatomy, duration of use and technology

02

Assess qualification, definitions and every applicable Annex VIII rule

03

Apply the strictest result and document the rationale for every condition

04

Determine the conformity-assessment route and scope of notified-body involvement

Controlled output

Class · rule · rationale · procedure

FDA

Device type, product code and premarket pathway

01

Define the intended use, indications for use, technology and key characteristics

02

Find the appropriate generic device type in the classification database and 21 CFR

03

Verify the product code, class, regulation number and limits of any exemption

04

Determine 510(k), De Novo, PMA or a justified exemption from premarket submission

Controlled output

Generic type · product code · class · pathway

Three devices, three different divergences

These cases illustrate the method and do not replace an assessment of the specific product. Even a small change in intended purpose or design can change the result.

Case 01

Reusable manual surgical drilling instrument

The instrument is used during surgery and is cleaned, disinfected and sterilised for reuse. It is not intended for connection to an active device.

EU

Class I under Rule 6

If the product meets the definition of a reusable surgical instrument, Rule 6 assigns class I. Notified-body involvement is limited to aspects of reuse, including cleaning, disinfection, sterilisation, maintenance and functional testing.

US

The exact generic device type is required

The phrase “surgical drill” is insufficient. The manufacturer must establish whether the product is a manual instrument, accessory, powered system or component, then verify the product code, regulation number, class and submission type.

What was corrected

Rule 6 applies to this type of device, not Rule 5. Its US class cannot be obtained by copying the European result.

A predicate does not classify the new device

Classification determines the legally defined device type and applicable controls. A predicate is used in a 510(k) to demonstrate substantial equivalence to a legally marketed device.

The search therefore begins with intended use, technology, product code and regulation number. An appropriate predicate is selected and justified only after confirming that the 510(k) route applies.

What the classification memorandum should preserve

A sound decision can be checked without an oral explanation from its author. It connects the product inputs to every regulatory conclusion.

01

Controlled product definition

Intended purpose, indications, users, environment, duration of contact, active nature, invasiveness, materials, software functions and variants.

02

EU rationale

Qualification, applicable definitions, rules considered, final class, exceptions and conformity-assessment route.

03

US rationale

Generic name, product code, regulation number, class, review panel, submission type, exemption limits and applicable controls.

04

Reassessment triggers

New claims or changes to the operating principle, materials, anatomy, duration, user, sterility or reusable status.

Classification is only the beginning

The US Quality Management System Regulation became effective on 2 February 2026 and incorporates ISO 13485:2016 by reference. A 510(k) exemption does not automatically exempt a device from every other FDA requirement.

In the EU, classification remains distinct from the clinical strategy. The 2026 amendments expanded the list of well-established technologies that may, when the conditions are met, be exempt from a mandatory clinical investigation. They did not remove clinical evaluation or change the device class by themselves.

Before fixing the class

Can an EU device class be converted into an FDA class?

No. The MDR and US legislation use different rules, categories and regulatory consequences. Each market requires a separate rationale based on one consistent definition of the device, its intended purpose and its characteristics.

Does a predicate device determine the US device class?

No. The device is first matched to an FDA generic device type, regulation number and product code. A predicate device is used later to support substantial equivalence when the 510(k) pathway applies.

Does every MDR class I device require a notified body?

No. Manufacturers generally self-declare ordinary class I devices. For sterile, measuring and reusable surgical class I devices, notified-body involvement is limited to the applicable special aspects.

Does 510(k)-exempt mean that no other FDA requirements apply?

No. The exemption concerns premarket notification and remains subject to applicable limitations. Establishment registration, device listing, labelling, reporting and QMSR requirements must be assessed separately.

When should the classification be established?

A working classification should be established before design requirements, the clinical strategy and product claims are finalised. It should be reassessed when the intended purpose, users, duration of use, technology or key characteristics change.

Fix the class and pathway for each market

MDRC reviews the intended purpose, technology and regulatory characteristics, prepares two separate rationales and connects the conclusions to the market-access plan.